Halloween to Cyber Monday sale: 30% off your first year of LMS Advisor, cloud or self-hosted.30% off your first year Ends Dec 1, 2026 Claim the offer
Compliance Training

QMSR training requirements: what changed for device makers in 2026

820.25 is gone. Since February 2, 2026, FDA inspects device training against ISO 13485 clause 6.2: competence, effectiveness, awareness and records of skills. Here is what to change.

QMSR training requirements cover illustration: a clipboard checklist for medical device competence and training records

The short answer: Since February 2, 2026, QMSR training requirements come from ISO 13485:2016 clause 6.2, which replaced the old 21 CFR 820.25. Device makers must define the competence each role needs, train or take other action, evaluate effectiveness in proportion to risk, make staff aware of how their work affects quality, and keep records of education, training, skills and experience in software they have validated for that use.

LMS Advisor handles the training and testing side of that list: role groups with assigned courses and due dates, scored knowledge checks and proctored exams as effectiveness evidence, and course certificates that expire for periodic retraining. It has no competency framework, no validation package and no Part 11 e-signature, so the controlled record often stays in your eQMS.

This guide is for quality managers, RA/QA consultants and training coordinators at US device makers updating their training procedure. We read the FDA pages and the final rule in September 2026. ISO 13485 is a paid standard, so where we quote clause text we name the secondary source and recommend you check it against your licensed copy.

What did the QMSR do to 820.25?

Effective date and ISO 13485 incorporation

FDA's QMSR page says the amended Part 820 "is effective February 2, 2026, and is now titled Quality Management System Regulation (QMSR)." It incorporates ISO 13485:2016 by reference and, in FDA's words, "harmonizes the FDA's CGMP regulatory framework with that used by other regulatory authorities." Under 21 CFR 820.10, a manufacturer must "Document a quality management system that complies with the applicable requirements of ISO 13485" as well as the other applicable requirements of Part 820.

The old training text vs clause 6.2

The current Part 820 has no separate training section. The former 820.25(b), as reproduced in the Kelsey Quality crosswalk, required each manufacturer to "establish procedures for identifying training needs" and to have all personnel trained to adequately perform their assigned responsibilities. It closed with four words: "Training shall be documented."

That was a training rule. Clause 6.2 is a competence rule. Training is one way to reach competence, and the clause adds duties the old text did not name: determining competence, checking effectiveness, building awareness, and recording skills and experience as well as courses.

FDA's reading in the final rule preamble

In the February 2, 2024 final rule, FDA answered a comment about qualified staff for verification and validation work. It said it "considers Clause 6.2 of ISO 13485 to capture the intent of the previous § 820.75(b)(1) adequately," and added: "All education, training, skills, and experience of employees need to be carefully recorded." That sentence is the one to show your team when someone argues that a course completion list is enough.

TopicBefore Feb 2, 2026 (820.25)From Feb 2, 2026 (ISO 13485 6.2 via QMSR)
Core dutyIdentify training needs, train personnel for their responsibilitiesDetermine necessary competence, then train or take other actions
EffectivenessNot stated in 820.25Evaluate effectiveness, proportionate to risk
AwarenessAwareness of device defects for relevant staffAwareness of relevance to quality objectives
RecordsTraining shall be documentedRecords of education, training, skills and experience
Software holding recordsGeneral validation expectationsClause 4.1.6 validation, proportionate to risk
Inspection approachQSITCompliance Program 7382.850

What does ISO 13485 clause 6.2 require, line by line?

As quoted by Greenlight Guru, clause 6.2 opens: "Personnel performing work affecting product quality shall be competent on the basis of appropriate education, training, skills and experience. The organization shall document the process(es) for establishing competence, providing needed training, and ensuring awareness of personnel." Then it lists five duties.

Determine competence

For each role whose work affects product quality, write down what competent looks like. A final inspector's profile might list the inspection SOPs, the defect catalog, gauge use and a minimum period of supervised inspection. This is a design document, and many 820.25-era procedures never wrote it down.

Train or take other actions

The clause, per the same Greenlight Guru article, says "provide training or take other actions to achieve or maintain the necessary competence." Other actions can mean hiring someone with the right experience, mentoring or assigning a qualified buddy. Training is the common answer, not the only one.

Evaluate effectiveness in proportion to risk

The organization must "evaluate the effectiveness of the actions taken" (quoted by both secondary sources), and a note to the clause, per the Kelsey Quality crosswalk, makes that check "proportionate to the risk associated with the work." A quiz may be enough for a document control SOP. A sterile barrier sealing operator needs an observed demonstration and a qualification record.

Awareness of quality objectives

Staff must know how their work affects quality and the quality objectives. Put a short, specific module in onboarding: what a field failure of this product looks like, and which step in their area prevents it.

Records of education, training, skills and experience

The clause, as Greenlight Guru quotes it, requires the organization to "maintain appropriate records of education, training, skills and experience." Skills and experience are the new part. A résumé summary, prior qualifications and supervised hours belong in the file next to course completions.

What can FDA inspectors see now?

QSIT retired, Compliance Program 7382.850 in use

On February 2, 2026, FDA "stopped using the Quality System Inspection Technique (QSIT)" and moved to the process in Compliance Program 7382.850. Expect inspectors to follow the ISO 13485 structure more closely, including competence and effectiveness.

Management review and audit reports are in scope

FDA's QMSR FAQ says the regulation lets FDA inspect management review, quality audit and supplier audit reports, because the old 820.180(c) exception is gone. If management review claims training is effective, the data behind that claim is now reviewable.

Records created before February 2, 2026

Investigators may review "records that are part of the manufacturer's QMS, including those created before February 2, 2026." Your legacy training files do not get a fresh start. The FAQ also says FDA "will not require certificates of conformance to ISO 13485," and a certificate "will not exempt a manufacturer from an FDA inspection."

Do you have to validate the software that holds training records?

Clause 4.1.6 in plain language

Yes, if the software holds QMS records. Clause 4.1.6, as quoted by Greenlight Guru, says such software "shall be validated prior to initial use and, as appropriate, after changes to such software or its application," and that the approach "shall be proportionate to the risk associated with the use of the software." Under the QMSR that applies in the US to any LMS, spreadsheet or eQMS that holds your QMS training records. The duty is yours, whatever the vendor says about its product.

CSA guidance, September 2025

FDA's Computer Software Assurance guidance gives risk-based recommendations for software "used as part of medical device production or the quality system." For a training system, that usually means testing the functions you rely on (assignment, completion, scoring, export) harder than the ones you do not. Our Part 11 and training records guide covers signatures and audit trails.

How do you update your training procedure for QMSR?

Competence profiles per role

Replace the "training matrix" section with competence profiles: required education or experience, training, effectiveness method, and requalification period for each role.

Effectiveness methods by risk

Assign one of three methods: a knowledge check for low-risk procedures, a scored exam for procedures with judgment calls, and an observed demonstration for operations where an error reaches the patient. The same risk logic appears in our GMP training requirements guide for drug makers.

Transition gap check on existing records

Pick ten people in quality-critical roles and pull their files. Can you show competence, not just completions, for each? Where you cannot, fix the gap now rather than during an inspection that may reach back before February 2026.

  1. Update the training SOP to cite ISO 13485 clause 6.2 instead of 820.25.
  2. Write competence requirements per role.
  3. Choose effectiveness methods by risk.
  4. Add quality awareness content to onboarding.
  5. Record skills and experience, not only courses.
  6. Check pre-2026 records for gaps inspectors may review.
  7. Validate, with a documented risk assessment, the software that holds training records.
  8. Include training effectiveness data in management review.

Keep the records for as long as your retention policy requires. Our training records retention guide covers the common US periods.

How LMS Advisor supports clause 6.2

LMS Advisor is our platform, so weigh that. Here is how it maps to the clause:

  • Determine competence and train. One group per role, organization course assignments with due dates and mandatory flags, and an overdue report.
  • Evaluate effectiveness. Scored knowledge checks in courses, and exams from a question bank with 14 question types. Image hotspot questions test component or defect identification. The Test Center adds native proctoring for qualification exams.
  • Demonstrations. Interactive video with questions at timestamps, and Recording Studio captures of a procedure. QR or manual attendance covers classroom sessions.
  • Periodic retraining. Course certificates carry a validity period and an expiring-soon status, and email automation flags people who never started.
  • Records and review. Audit logs, CSV, Excel or PDF exports for management review, and webhooks or the REST API to your eQMS.

Where a specialist QMS is the better fit: LMS Advisor has no competency framework or skills matrix, no validation package, no Part 11 e-signature and no retraining triggered by document control. As of September 2026, Greenlight Guru says training "can be automatically triggered by new or revised SOPs and QMS documents," and Qualio says "affected training assignments update automatically" when a record changes. If that is your main need, use one of them and consider LMS Advisor for customer, distributor and rep training. Our medical device training software guide and the pharmaceutical and life sciences page cover that split. To see it with your own procedures, book a demo.

Frequently asked questions

Does QMSR still require training records?

Yes. ISO 13485 clause 6.2, enforceable through the QMSR, requires appropriate records of education, training, skills and experience. FDA's final rule adds that these "need to be carefully recorded."

What replaced 21 CFR 820.25?

ISO 13485:2016 clause 6.2, incorporated by reference into Part 820 from February 2, 2026. The current Part 820 no longer has its own training section.

Does ISO 13485 require evaluating training effectiveness?

Yes. Clause 6.2 requires the organization to evaluate the effectiveness of the actions taken, with the method proportionate to the risk of the work. Scored tests, observed demonstrations and trend reviews are common evidence.

Do we need ISO 13485 certification to comply with QMSR?

No. FDA's QMSR FAQ says it will not require certificates of conformance to ISO 13485, and a certificate does not exempt you from inspection.

Does our LMS need to be validated under QMSR?

If it holds QMS records, yes. Clause 4.1.6 requires validating the use of QMS software before first use and after changes, proportionate to risk. The manufacturer owns that validation.

Can LMS Advisor support QMSR training requirements?

Yes, for delivery and effectiveness evidence: role assignments, scored checks, proctored exams and expiring course certificates, with exports to your eQMS. It is not validated for you and has no competency framework or Part 11 e-signature, so you validate your use or keep controlled records in a validated eQMS.

Add LMS Advisor to your preferred sources

Add lmsadvisor.com as a preferred source on Google, or open this article in your AI assistant to use it as a source.

Swati Priyadarshani
· Founder & CEO at LMS Advisor

I'm Swati Priyadarshani, Founder & CEO of LMS Advisor and Co-Founder of WorldWin Coder Pvt. Ltd. (est. 2019). Over the last 10+ years, I've helped enterprises build learning platforms that actually get used - not just deployed.

Keep reading

See LMS Advisor with your own use case

A product specialist walks you through authoring, a proctored exam, certificates and the AI tools, using the programs you actually run.